Accuracy evaluation of time domain measurement systems for microwave tomography
Paper i proceeding, 2009

The measurement accuracy of time domain systems for microwave tomography are evaluated through system level simulations and compared with that of a network analyzer based frequency domain system. In the analysis, both a time domain ultra-wideband (UWB) impulse and multiband system architectures are considered. It is shown that the measurement error of the time domain systems mainly comes from analog to digital (A/D) conversion. For a simple comparison case, the relative measurement error involved in an estimated object response signal from the time domain UWB impulse system and UWB multiband system measurements are around three times and ten times higher than that from frequency domain system.

system modeling

medical imaging

time domain measurements.

Measurement errors

microwaveimaging

Författare

Xuezhi Zeng

Chalmers, Signaler och system, Signalbehandling och medicinsk teknik

Andreas Fhager

Chalmers, Signaler och system, Signalbehandling och medicinsk teknik

Peter Linner

Chalmers, Mikroteknologi och nanovetenskap, Mikrovågselektronik

Mikael Persson

Chalmers, Signaler och system, Signalbehandling och medicinsk teknik

Herbert Zirath

Chalmers, Mikroteknologi och nanovetenskap, Mikrovågselektronik

IEEE MTT-S International Microwave Symposium Digest

0149645X (ISSN)

1441-1444 5165978
978-142442804-5 (ISBN)

Ämneskategorier

Annan elektroteknik och elektronik

DOI

10.1109/MWSYM.2009.5165978

ISBN

978-142442804-5

Mer information

Skapat

2017-10-08