Convex optimization of sensor positions for organ-positioning during radiotherapy
Other conference contribution, 2012

Real-time organ positioning is a promising technique for enhancing radiotherapy of cancer tumors in situations where the tumor position can be believed to change between imaging and treatment and/or during treatment. In this work, we exploit theory on optimal measurements to formulate an optimization problem for the sensor positions of a magnetic tracking system. The optimization problem is relaxed into a convex problem, which is subsequently solved.

optimal measurements

prostate cancer

magnetic tracking

radiotherapy

real-time organ-positioning

Author

Oskar Talcoth

Chalmers, Signals and Systems, Signal Processing and Biomedical Engineering

Thomas Rylander

Chalmers, Signals and Systems, Signal Processing and Biomedical Engineering

Proceedings of Medicinteknikdagarna 2012

87-

Areas of Advance

Life Science Engineering (2010-2018)

Subject Categories

Electrical Engineering, Electronic Engineering, Information Engineering