A biocompatible micro cell culture chamber (microCCC) for the culturing and on-line monitoring of eukaryote cells
Journal article, 2006

We have previously shown that a polymeric (PMMA) chip with medium perfusion and integrated heat regulation provides sufficiently precise heat regulation, pH-control and medium exchange to support cell growth for weeks. However, it was unclear how closely the cells cultured in the chip resembled cells cultured in the culture flask. In the current study, gene expression profiles of cells cultured in the chip were compared with gene expression profiles of cells cultured in culture flasks. The results showed that there were only two genes that were differently expressed in cells grown in the cell culture chip compared to cell culture flasks. The cell culture chip could without further modification support cell growth of two other cell lines. Light coming from the microscope lamp during optical recordings of the cells was the only external factor identified, that could have a negative effect on cell survival. Low grade light exposure was however compatible with optical recordings as well as cell viability. These results strongly indicate that a cell culture chip could be constructed that allowed for on-line optical recording of cellular events without affecting the cell culturing condition compared to cell cultured in culture flasks incubated in a dark and CO 2 conditioned incubator. © The Royal Society of Chemistry 2006.

Author

M. Stangegaard

Technical University of Denmark (DTU)

Sarunas Petronis

Chalmers, Applied Physics, Chemical Physics

A.M. Jorgensen

Technical University of Denmark (DTU)

C.B. Christensen

Coloplast Danmark AS

M. Dufva

Technical University of Denmark (DTU)

Lab on a Chip - Miniaturisation for Chemistry and Biology

1473-0197 (ISSN) 1473-0189 (eISSN)

Vol. 6 8 1045-1051

Subject Categories

Other Engineering and Technologies

DOI

10.1039/b603379b

More information

Latest update

2/28/2018