Evaluation of novel radiation protection devices during radiologically guided interventions
Journal article, 2024

Background:
In radiologically guided interventions, medical practitioners are subjected to radiation exposure, which may lead to radiation-induced diseases. In this study, novel radiation shields for the head and neck were evaluated for their potential to reduce radiation exposure.
Method:
An anthropomorphic phantom was exposed on its left side to scattered radiation from beneath to simulate the exposure of an operator in a x-ray operating room. Thermoluminescent dosimeters (TLDs) were positioned at different depths in five slices in the phantom, measuring personal dose equivalent. Two different set up situations were evaluated: a head protector designed to reduce radiation in the upper section of the head; and a novel thyroid protector prototype extended in the front and on both sides, designed to reduce radiation in the lower and middle sections of the head. A standard thyroid collar prototype and a ceiling mounted lead glass shield were used as comparisons. Furthermore, the head protector was evaluated in a clinical study in which TLDs were positioned to measure scattered radiation exposure to the heads of operators during endovascular interventions.
Results:
The extended thyroid protector reduced the scattered radiation in the throat, chin, and ear slices. Some shielding effect was seen in the brain and skull slices. The head protector showed a shielding effect in the skull slice up to two cm depth where it covered the phantom head. As expected, the ceiling mounted lead glass shield reduced the scattered radiation in all measuring points.
Conclusions:
A ceiling mounted lead glass shield is an effective radiation protection for the head, but in clinical practice, optimal positioning of a ceiling mounted lead shield may not always be possible, particularly during complex cases when radiation protection may be most relevant. Added protection using these novel guards may compliment the shielding effect of the ceiling mounted lead shield. The head protector stand-alone did not provide sufficient protection of the head. The extended thyroid protector stand-alone provided sufficient protection in the lower and middle sections of the head and neck. Graphical Abstract: (Figure presented.)

Radiation exposure

Scatter radiation

Ionizing radiation

Personal protection equipment

Radiation protection

PPE

Author

Maria E.V. Larsson

Sahlgrenska University Hospital

University of Gothenburg

Pernilla I. Jonasson

University of Gothenburg

Petra Apell

Texray AB

Chalmers, Technology Management and Economics, Innovation and R&D Management

Peter P. Kearney

University College Cork

Charlotta J. Lundh

Sahlgrenska University Hospital

University of Gothenburg

CVIR Endovascular

25208934 (eISSN)

Vol. 7 1 18

Subject Categories

Radiology, Nuclear Medicine and Medical Imaging

DOI

10.1186/s42155-024-00430-0

More information

Latest update

3/6/2024 3