Toward enhanced quality assurance guidelines for deep hyperthermia devices: a multi-institution study
Journal article, 2024

Introduction: Hyperthermia efficacy depends on the temperatures achieved in the target area. Therefore, hyperthermia systems must deliver both controlled and conformal heating. This study presents a comprehensive multi-institutional quality assurance (QA) evaluation of deep hyperthermia devices. Methods: Six European institutions equipped with BSD- Sigma 60 and Sigma Eye deep hyperthermia applicators participated in the study. Up to six measurements per applicator were performed in each institution. The thermal distribution in cylindrical homogeneous phantoms after 10 minutes of heating with a total power delivered of 1000 watts was assessed using the applicator’s integrated mapping thermometry system. Evaluated quality parameters included temperature increase, focus location, and focus symmetry. Results: A total of 54 measurements were conducted, with 43 included in the analysis. All applicators, except one, achieved a temperature increase of 6 °C in 10 minutes. Central heating capabilities were demonstrated, with mean deviations from the intended location of −1.4 ± 1.6 cm for Sigma 60 and 1.5 ± 1.4 cm for Sigma Eye. Symmetry evaluations showed differences in radial temperature profiles of 6.2 ± 4.5 % for the Sigma 60 and 5.9 ± 4.4 % for the Sigma Eye. We propose minimum acceptable values for each quality parameter based on these results. Conclusion: The measurements were reproducible with acceptable values for the various quality parameters. Potential deviations might be attributed to inaccuracies in the mapping thermometry system rather than the heating system. The presented protocol and practical recommendations should be applied for future QA measurements in deep hyperthermia.

homogeneous phantom

quality assurance

measurement protocol

deep hyperthermia applicators

Hyperthermia

Author

Mattia de Lazzari

Chalmers, Electrical Engineering, Signal Processing and Biomedical Engineering

Carolina Carrapiço-Seabra

Erasmus MC Cancer Institute

D. Marder

Kantonsspital Aarau

G. C. Van Rhoon

Erasmus MC Cancer Institute

Delft University of Technology

S. Curto

Erasmus MC Cancer Institute

Hana Dobsicek Trefna

Chalmers, Electrical Engineering, Signal Processing and Biomedical Engineering

International Journal of Hyperthermia

0265-6736 (ISSN) 1464-5157 (eISSN)

Vol. 41 1 2436005

Creation of advanced cancer treatment planning to boost the effect of Radiotherapy by combining with hyperthermia, heating the tumor (HYPERBOOST)

European Commission (EC) (EC/H2020/955625), 2020-12-01 -- 2024-11-30.

Subject Categories

Energy Engineering

DOI

10.1080/02656736.2024.2436005

More information

Latest update

12/20/2024