Study to Achieve Stable VNA Measurements Using 500-750 GHz Quasi-Optical Setup
Paper in proceeding, 2026

A quasi-optical S-parameter measurement system in the 500–750 GHz (WM-380) frequency range for high-loss samples was characterized for its dynamic range and stability using Allan deviation. The system provides a dynamic range of more than 114 dB in the 630–690 GHz frequency range. Allan deviation was used to find the highest stability and sensitivity, as a trade-off between mitigating noise and drift, by finding the right combinations of intermediate frequency (IF) bandwidth of the vector network analyzer and number of averaged measurements. We observe that a 600-μm-thick water sample, measured with an IF bandwidth of 100 Hz with 16 averages yields a stability better than 0.01 dB.

THz

high-loss samples

quasi-optics

water

VNA

Allan deviation

Author

Daniel Georg Hellmich

Chalmers, Microtechnology and Nanoscience (MC2), Terahertz and Millimetre Wave Laboratory

Kasturie Dileep Jatkar

Chalmers, Microtechnology and Nanoscience (MC2), Terahertz and Millimetre Wave Laboratory

Helena Rodilla

Chalmers, Microtechnology and Nanoscience (MC2), Terahertz and Millimetre Wave Laboratory

56th European Microwave Conference, EuMC 2026

56th European Microwave Conference, EuMC 2026
London, United Kingdom,

Unveiling terahertz dynamics of biomacromolecules in solutions

Swedish Research Council (VR) (2024-01981), 2025-01-01 -- 2029-12-31.

Infrastructure

Kollberg Laboratory

Areas of Advance

Nanoscience and Nanotechnology

Subject Categories (SSIF 2025)

Other Electrical Engineering, Electronic Engineering, Information Engineering

DOI

10.23919/EuMC69230.2026.11748148

More information

Created

11/13/2026