An Oscilloscope Correction Method for Vector-Corrected RF Measurements
Journal article, 2015

The transfer characteristics of the RF front-end circuitry of a real-time oscilloscope (RTO) are not only frequency dependent and nonlinear with signal amplitude but also dependent on the voltage range setting of the oscilloscope. A correction table in the frequency domain is proposed to account for the additional gain and delay introduced when switching between different voltage ranges. The table was extracted from the measured data in which a continuous-wave signal source was connected to the oscilloscope ports, whereas the input power, the frequency, and voltage ranges were varied. The importance of the corrections is demonstrated by its use in an RTO-based two-port vector-corrected measurement system. Measurements from the oscilloscope-based system are compared with a vector network analyzer (VNA), leading to less discrepancy in the measured Sparameters of two amplifiers when using the proposed correction technique.

frequency-domain analysis

microwave measurements

dynamic range

Calibration

oscilloscopes

Author

Sebastian Gustafsson

Chalmers, Microtechnology and Nanoscience (MC2), Microwave Electronics

GigaHertz Centre

Mattias Thorsell

Chalmers, Microtechnology and Nanoscience (MC2), Microwave Electronics

GigaHertz Centre

Jörgen Stenarson

Chalmers, Microtechnology and Nanoscience (MC2), Microwave Electronics

SP Sveriges Tekniska Forskningsinstitut AB

Christian Fager

GigaHertz Centre

Chalmers, Microtechnology and Nanoscience (MC2), Microwave Electronics

IEEE Transactions on Instrumentation and Measurement

0018-9456 (ISSN) 1557-9662 (eISSN)

Vol. 64 9 2541-2547 7058369

Areas of Advance

Information and Communication Technology

Subject Categories

Signal Processing

DOI

10.1109/tim.2015.2407451

More information

Latest update

4/5/2022 6