Development of a High-Tc SQUID-Based System for Neurophysiology Studies In-Vitro
Paper in proceeding, 2006

In this paper we report on the development of a system based on a high-T c SQUID (HTS) sensor for measurements of the neuromagnetic field generated by neurons inside tissue slices. SQUIDs have successfully been measured inside the system. The system white noise level is lower than 7 pT/Hz 1/2 , which is only slightly higher than previously reported 4.5 pT/Hz 1/2 for the same kind of SQUID measured inside a superconducting shield. © 2006 IOP Publishing Ltd.

Author

Per Magnelind

Chalmers, Microtechnology and Nanoscience (MC2), Quantum Device Physics

E. Tarte

University of Birmingham

Dag Winkler

Chalmers, Microtechnology and Nanoscience (MC2), Quantum Device Physics

Eric Hanse

University of Gothenburg

Journal of Physics: Conference Series

1742-6596 (eISSN)

Vol. 43 1 1243-1246

Subject Categories

Physical Sciences

DOI

10.1088/1742-6596/43/1/303