Development of a High-Tc SQUID-Based System for Neurophysiology Studies In-Vitro
Paper i proceeding, 2006

In this paper we report on the development of a system based on a high-T c SQUID (HTS) sensor for measurements of the neuromagnetic field generated by neurons inside tissue slices. SQUIDs have successfully been measured inside the system. The system white noise level is lower than 7 pT/Hz 1/2 , which is only slightly higher than previously reported 4.5 pT/Hz 1/2 for the same kind of SQUID measured inside a superconducting shield. © 2006 IOP Publishing Ltd.

Författare

Per Magnelind

Chalmers, Mikroteknologi och nanovetenskap, Kvantkomponentfysik

E. Tarte

University of Birmingham

Dag Winkler

Chalmers, Mikroteknologi och nanovetenskap, Kvantkomponentfysik

Eric Hanse

Göteborgs universitet

Journal of Physics: Conference Series

1742-6596 (eISSN)

Vol. 43 1 1243-1246

Ämneskategorier

Fysik

DOI

10.1088/1742-6596/43/1/303

Mer information

Senast uppdaterat

2018-02-28