Experimental validation of corrections factors for gamma-gamma and gamma-X coincidence summing of Ba-133, Eu-152, and Sb-125 in volume sources
Journal article, 2020

True coincidence summing correction factors for Ba-133, Eu-152 and Sb-125 were determined experimentally for a small volume source and compared with correction factors obtained with three softwares (EFFTRAN-X, GESPECOR and VGSL). The radionuclides investigated have a relatively challenging decay scheme and their spectra are known to suffer from losses due to summation (gamma-gamma, gamma-X and X-X) when measured at close distances on a HPGe detector sensitive to low energy photons. This study shows that the softwares were in good agreement with each other and the experimental data and the calculated activity was consistent with the activity in the volume source.

Sb-125

EFFTRAN-X

VGSL

Ba-133

GESPECOR

Coincidence summing

Eu-152

Author

S. Jonsson

Swedish Defence Research Agency (FOI)

J. Kastlander

Swedish Defence Research Agency (FOI)

T. Vidmar

Belgian Nuclear Research Center (SCK CEN)

Henrik Ramebäck

Chalmers, Chemistry and Chemical Engineering

Swedish Defence Research Agency (FOI)

Journal of Radioanalytical and Nuclear Chemistry

0236-5731 (ISSN) 1588-2780 (eISSN)

Vol. 323 1 465-472

Subject Categories

Accelerator Physics and Instrumentation

Subatomic Physics

Astronomy, Astrophysics and Cosmology

DOI

10.1007/s10967-019-06938-3

More information

Latest update

3/18/2021