Estimating Analytical Errors of Glomerular Filtration Rate Measurement
Journal article, 2022

Background Few studies are available on how to optimize time points for sampling and how to estimate effects of analytical uncertainty when glomerular filtration rate (GFR) is calculated. Methods We explored the underlying regression mathematics of how analytical variation of a kidney filtration marker affects 1-compartment, slope-and-intercept GFR calculations, using 2 or 3 time points following a bolus injection, and used this to examine the results from 731 routine 3-point iohexol plasma clearance measurements. Results GFR calculations inflated analytical uncertainty if the time points were taken too late after the bolus injection and too close after each other. The uncertainty in GFR calculation was, however, the same as the analytical uncertainty if optimal time points were used. The middle of the 3 samples was of little value. The first sample should be taken as early as possible after the distribution phase. Sampling before the patient specific half-life of the kidney filtration marker resulted in an exponential error inflation whereas no error inflation was seen when sampling occurred later than 2 half-lives. Theoretical GFR uncertainty could be lowered 2.6-fold if individually optimized time points for sampling had been used in our 731 clearance measurements. Using Taylor expansions to approximate the moments of transformed random variables, the uncertainty of an individual GFR measurement could be calculated in a simple enough way to be applicable by laboratory software. Conclusions We provide a theoretical foundation to select patient-optimal time points that may both limit errors and allow calculation of GFR uncertainty.

Author

Damiano Ognissanti

Chalmers, Mathematical Sciences, Algebra and geometry

University of Gothenburg

Moa Andresen Bergstrom

University of Gothenburg

Sahlgrenska University Hospital

Elvar Theodorsson

Linköping University

Anders Larsson

Uppsala University

Gunnar Nordin

EQUALIS

Ola Hammarsten

Sahlgrenska University Hospital

University of Gothenburg

Clinical Chemistry

0009-9147 (ISSN) 1530-8561 (eISSN)

Vol. 68 9 1211-1218

Subject Categories

Medical Laboratory and Measurements Technologies

Analytical Chemistry

Probability Theory and Statistics

DOI

10.1093/clinchem/hvac098

PubMed

35802029

More information

Latest update

9/28/2022