A planar picoamperemeter based on a superconducting quantum interferometer
Artikel i vetenskaplig tidskrift, 2006

An optimized picoamperemeter based on a superconducting quantum interferometer device (SQUID) for a metal cold-electron bolometer is fabricated and experimentally studied. The intrinsic SQUID current noise caused by the input coil is estimated to be less than 1 pA/Hz@1/2. Owing to the application of modulation electronics, current sensitivity in the input coil reaches 5 pA/Hz@1/2 in the frequency band from 10 Hz to 10 kHz.

SQUID

interferometer

bolometer

Readout

Författare

Eugeny Burmistrov

Chalmers, Mikroteknologi och nanovetenskap, Kvantkomponentfysik

Mikhail Tarasov

Chalmers, Mikroteknologi och nanovetenskap, Kvantkomponentfysik

Leonid Kuzmin

Chalmers, Mikroteknologi och nanovetenskap, Kvantkomponentfysik

J. Communications Technol. and Electron.,

Vol. 51 11 1319-1324

Ämneskategorier

Den kondenserade materiens fysik

Mer information

Skapat

2017-10-07