Intra-observer repeatability when assessing the foetal urinary bladder volume by the Virtual Organ Computer-aided AnaLysis and SUM-OF-CYLINDERS methods: A pilot study
Artikel i vetenskaplig tidskrift, 2017
Method: The urinary bladders of 20 foetuses were documented by three-dimensional ultrasound. Standard deviation was compared when the volumes of identical bladder images were repeatedly estimated using the Virtual Organ Computer-aided AnaLysis and the experimental SUM-OF-CYLINDERS methods.
Results: No systematic deviation was found between the estimated volumes when using these two methods. Standard deviation was smaller for the SUM-OF-CYLINDERS compared to the Virtual Organ Computer-aided AnaLysis method (p < 0.0001). In relation to bladder volumes of 5–25 ml, standard deviation was 11–14% for the Virtual Organ Computer-aided AnaLysis and 4–5% for the SUM-OF-CYLINDERS method.
Conclusions: Using three-dimensional ultrasound images adapted for the Virtual Organ Computer-aided AnaLysis method, foetal urinary bladder volumes can also be estimated using the SUM-OF-CYLINDERS method. The SUM-OF-CYLINDERS method employs technical advances which may result in a lower standard deviation and therefore higher intra-observer repeatability.
ultrasonography
diagnostic technique and procedure
Foetus
urinary bladder
reliability
Författare
M. Fagerquist
NU-sjukvården
Innlandet Hospital Trust
Vivek Sethi
Akershus University Hospital
E. Skytteren
Innlandet Hospital Trust
Anders Odén
Chalmers, Matematiska vetenskaper, Tillämpad matematik och statistik
Ultrasound
1742-271X (ISSN) 17431344 (eISSN)
Vol. 25 3 156-165Ämneskategorier
Radiologi och bildbehandling
Datorseende och robotik (autonoma system)
Medicinsk bildbehandling
DOI
10.1177/1742271X17709854