An Inter-Laboratory Comparison of NVNA Measurements
Paper i proceeding, 2018

A comparison of nonlinear vector network analyser (NVNA) measurements has been carried out involving four organisations (National Physical Laboratory, UK, University of Surrey, UK, Chalmers University of Technology, Sweden and Keysight Technologies, Denmark). Three nonlinear devices consisting of two amplifiers and a nonlinear verification device (NLVD) were measured by each of the organisations. Results are presented which show generally good agreement between the measurements and give some indication of the typical amount of variability to be expected in measurements of this type.

Nonlinear microwave measurements

NVNA

measurement comparison

Författare

Martin Salter

National Physical Laboratory (NPL)

Laurence Stant

University of Surrey

Koen Buisman

Chalmers, Mikroteknologi och nanovetenskap, Mikrovågselektronik

Troels Nielsen

Keysight Technologies

International Workshop on Integrated Nonlinear Microwave and Millimetre-Wave Circuits, INMMIC 2018 - Proceedings

8430012

2018 International Workshop on Integrated Nonlinear Microwave and Millimetre-Wave Circuits, INMMIC 2018
Brive La Gaillarde, France,

Ämneskategorier

Acceleratorfysik och instrumentering

Annan fysik

Marin teknik

DOI

10.1109/INMMIC.2018.8430012

Mer information

Senast uppdaterat

2019-01-17