Performance Evaluation of a Time Domain Microwave System for Medical Diagnostics
Artikel i vetenskaplig tidskrift, 2019

A time-domain microwave system for medical diagnostics was developed using off-the-shelf components. The system's performance was evaluated for realistic measurement scenarios when two different measurement strategies: with and without variable gain controls (VGCs) were employed. The results showed that without the employment of the VGC strategy, the system was able to measure loss up to 80 dB in the frequency range of 500 MHz-3 GHz. A 20 dB higher dynamic range was obtained when the VGC strategy was used. This measurement accuracy was obtained with settings which results in the measurement speed of 20 μs per trace.

Författare

Xuezhi Zeng

Chalmers, Elektroteknik, Signalbehandling och medicinsk teknik

IEEE Transactions on Instrumentation and Measurement

0018-9456 (ISSN) 1557-9662 (eISSN)

Vol. 68 8 2880-2889 8472295

Ämneskategorier

Elektroteknik och elektronik

DOI

10.1109/TIM.2018.2867976

Mer information

Senast uppdaterat

2022-04-05