Background and Objective: Commercially available motorized prosthetic legs use exclusively non-biological signals to control movements, such as those provided by load cells, pressure sensors, and inertial measurement units (IMUs). Despite that the use of biological signals of neuromuscular origin can provide more natural control of leg prostheses, these signals cannot yet be captured and decoded reliably enough to be used in daily life. Indeed, decoding motor intention from bioelectric signals obtained from the residual limb holds great potential, and therefore the study of decoding algorithms has increased in the past years with standardized methods yet to be established.

Methods: In the absence of shared tools to record and process lower limb bioelectric signals, such as electromyography (EMG), we developed an open-source software platform to unify the recording and processing (pre-processing, feature extraction, and classification) of EMG and non-biological signals amongst researchers with the goal of investigating and benchmarking control algorithms. We validated our locomotion decoding (LocoD) software by comparing the accuracy in the classification of locomotion mode using three different combinations of sensors (1 = IMU+EMG, 2 = EMG, 3 = IMU). EMG and non-biological signals (from the IMU and pressure sensor) were recorded while able-bodied participants (n = 21) walked on different surfaces such as stairs and ramps, and this data set is also released publicly along this publication. LocoD was used for all recording, pre-processing, feature extraction, and classification of the recorded signals. We tested the statistical hypothesis that there was a difference in predicted locomotion mode accuracy between sensor combinations using the Wilcoxon signed-rank test.

Results: We found that the sensor combination 1 (EMG+IMU) led to significantly more accurate and improved locomotion mode prediction (Accuracy=93.4 ± 3.9) than using EMG (Accuracy= 74.56 ± 5.8) or IMU alone (Accuracy=90.77 ± 4.6) with p-value < 0.001.

Conclusions: Our results support previous research and validate the functionality of LocoD as an open-source and modular platform to research control algorithms for prosthetic legs that incorporate bioelectric signals.

Författare

Bahareh Ahkami

Chalmers, Elektroteknik, System- och reglerteknik

Kirstin Ahmed

Chalmers, Elektroteknik, System- och reglerteknik

Morten Kristoffersen

Chalmers, Elektroteknik, System- och reglerteknik

Max Jair Ortiz Catalan

Chalmers, Elektroteknik, System- och reglerteknik

Computer Methods and Programs in Biomedicine

0169-2607 (ISSN) 18727565 (eISSN)

Ämneskategorier

Elektroteknik och elektronik

DOI

10.2139/ssrn.4575926

Mer information

Skapat

2024-02-13