Development of an integrated Face-Upper airway model
Övrigt konferensbidrag, 2025
This study aims to demonstrate the influence of a realistic infant airway model on the delivery of a respirable aerosol for improved understanding and to better predict drug delivery to the lungs.
To further expand the in-vitro characterization capabilities for paediatric development programs, an advanced upper airway model, “Frank”, representing a 2-year-old child has been developed. The model is originally based on scans of a 35-year-old male which has been scaled to represent approx. a 2-year-old child using published, tabular data.
The aerosol filtering effect of the Frank model has been investigated at flow rates representative of the age group (3.5 and 7 L/min) and using a soft mist inhaler.
A 50% reduction of the aerosol dose passing the airway model compared to reference runs without model, was found.
aerosol filtering effect