Room Temperature Interface for RSFQ Digital Signal Processor
Artikel i vetenskaplig tidskrift, 2007

This paper describes a cryogenic Multi-Chip Module (MCM) probe for testing of a superconducting hybrid Digital Signal Processor (DSP) capable of 24 giga multiply accumulate operations per second. The probe has 39 data channels with 4 GHz bandwidth, a 30 GHz high speed clock input line, and 48 biasing lines. The MCM interface board features 164 spring finger contacts for a 30 mm x 30 mm MCM. The probe has low heatload (0.23 W) and provides good magnetic shielding with residual field equal to 250 Flux quanta over a 5 mm x 5 mm area. The measured time skew between 20 of the 1200 mm data coaxial lines is about 70 ps. The probe allows synchronous data word transmission at 2 Gbps.

Cryoprobe MCM RSFQ heatloads magnetic shielding

Författare

Henrik Engseth

Chalmers, Mikroteknologi och nanovetenskap, Kvantkomponentfysik

Raihan Rafique

Chalmers, Mikroteknologi och nanovetenskap, Mikrovågselektronik

Irina Kataeva

Chalmers, Mikroteknologi och nanovetenskap, Kvantkomponentfysik

Samuel Intiso

Chalmers, Mikroteknologi och nanovetenskap

Anna Kidiyarova-Shevchenko

Chalmers, Mikroteknologi och nanovetenskap

IEEE Transactions on Applied Superconductivity

Vol. 17 2 979-982

Ämneskategorier

Fysik

Elektroteknik och elektronik

Mer information