Equivalence of computer codes for calculation of coincidence summing correction factors - Part II
Journal article, 2016

The aim of this study was to check for equivalence of computer codes that are capable of performing calculations of true coincidence summing (TCS) correction factors. All calculations were performed for a set of well-defined detector parameters, sample parameters and decay scheme data. The studied geometry was a point source of Ba-133 positioned directly on the detector window of a low-energy (n-type) detector. Good agreement was established between the TCS correction factors computed by the different codes.

True coincidence summing

Computer codes

Gamma-ray spectrometry

High-purity germanium detector

Intercomparison

Author

T. Vidmar

Belgian Nuclear Research Center (SCK CEN)

A. Camp

Polytechnic University of Catalonia

S. Hurtado

University of Seville

H. Jaderstrom

Canberra Industries, Inc.

J. Kastlander

Swedish Defence Research Agency (FOI)

M. C. Lepy

The French Alternative Energies and Atomic Energy Commission (CEA)

G. Lutter

European Commission (EC)

Henrik Ramebäck

Chalmers, Chemistry and Chemical Engineering, Energy and Material

O. Sima

Universitatea din Bucuresti

A. Vargas

Polytechnic University of Catalonia

Applied Radiation and Isotopes

0969-8043 (ISSN) 18729800 (eISSN)

Vol. 109 482-486

Subject Categories

Atom and Molecular Physics and Optics

DOI

10.1016/j.apradiso.2015.11.071