Equivalence of computer codes for calculation of coincidence summing correction factors - Part II
Artikel i vetenskaplig tidskrift, 2016

The aim of this study was to check for equivalence of computer codes that are capable of performing calculations of true coincidence summing (TCS) correction factors. All calculations were performed for a set of well-defined detector parameters, sample parameters and decay scheme data. The studied geometry was a point source of Ba-133 positioned directly on the detector window of a low-energy (n-type) detector. Good agreement was established between the TCS correction factors computed by the different codes.

True coincidence summing

Computer codes

Gamma-ray spectrometry

High-purity germanium detector

Intercomparison