Noise performance comparison between two different types of time-domain systems for microwave detection
Journal article, 2017

This paper compares the noise performance of two different types of time-domain microwave detection systems: a pulsed system and a pseudo-random noise sequence system. System-level simulations and laboratory-based measurements are carried out in the study. Results show that the effect of timing jitter is more significant on the measurement accuracy of the pseudo-random noise sequence system than that of the pulsed system. Although the signal power density of the pseudo-random sequence system is tens of dBs higher than that of the pulsed system over the frequency band of interest, the signal-to-noise ratio difference between these two systems can be just a few dBs or even smaller depending on the jitter level.

Quality of life/medical diagnosis and imaging systems

Microwave measurements

Author

Xuezhi Zeng

Chalmers, Signals and Systems, Signal Processing and Biomedical Engineering

Albert Monteith

Chalmers, Earth and Space Sciences, Radar Remote Sensing

Andreas Fhager

Chalmers, Signals and Systems, Signal Processing and Biomedical Engineering

Mikael Persson

Chalmers, Signals and Systems, Signal Processing and Biomedical Engineering

Herbert Zirath

Chalmers, Microtechnology and Nanoscience (MC2), Microwave Electronics

International Journal of Microwave and Wireless Technologies

1759-0787 (ISSN) 1759-0795 (eISSN)

Vol. 9 3 535-542

Subject Categories

Medical Equipment Engineering

Signal Processing

Areas of Advance

Life Science Engineering (2010-2018)

DOI

10.1017/S175907871600057X