Impact of different headgear on release of bacteria when wearing a clean air suit for surgery: an experimental study
Journal article, 2026
Methods: Tests with five healthy males wearing a clean air suit and different types of headgear were performed in a dispersal chamber according to Annex E in European standard EN 13795-2:2019.
Results: The mean source strength value (number of bacteria emitted per second from one person) was 7.3 CFU/s without headgear and 7.2 CFU/S with a surgical cap (P = 0.52). The mean source strength value obtained with a disposable surgical hood was 5.3 CFU/s, which was not significantly lower than that obtained without headgear (P = 0.057).
Discussion: The poor performance of the disposable surgical headgear was likely an effect of both permeable material and design. The mean source strength value with a textile hood in the previous study was 1.0 CFU/s, which was significantly lower than the mean source strength value without headgear (P < 0.01). Conclusion: When high microbial cleanliness of the air in the operating room is required, staff should wear a clean air suit fulfilling the criteria in European standard EN 13795-2, and a hood made of the same material as the rest of the suit.
Disposable hood
Microbial air cleanliness
Reusable hood
Headgear
Disposable cap
Author
Bengt Ljungqvist
Chalmers, Architecture and Civil Engineering, Building Services Engineering
Berit Reinmüller
Chalmers, Architecture and Civil Engineering, Building Services Engineering
Ann Tammelin
Karolinska Institutet
Infection Prevention in Practice
25900889 (eISSN)
Vol. 8 2 100530Subject Categories (SSIF 2025)
Building Technologies
DOI
10.1016/j.infpip.2026.100530