Calculating the Current Density and Electric Field in Human Head by Multichannel Transcranial Magnetic Stimulation
Artikel i vetenskaplig tidskrift, 2009

In this paper, we present a designed multichannel transcranial magnetic stimulation (mTMS) system with 40 small coils. The induced current density and electric field in realistic human head model by mTMS were calculated by impedance method and the results were compared with figure-of-eight coil. The dependence of stimulation focality on the coil current direction was investigated. Properly adjusting the input current phases can improve the magnitude of the induced fields in the brain.

coils

impedance method

human motor cortex

multichannel

Current density

tissues

electric field

brain

transcranial magnetic stimulation (mTMS)

Författare

M. Lu

Faculdade de Ciencias, Universidade de Lisboa

S. Ueno

Kyushu University

T. Thorlin

Sahlgrenska universitetssjukhuset

Mikael Persson

Chalmers, Signaler och system, Signalbehandling och medicinsk teknik, Biomedicinsk elektromagnetik

IEEE Transactions on Magnetics

0018-9464 (ISSN)

Vol. 45 3 1662-1665

Ämneskategorier

Elektroteknik och elektronik

DOI

10.1109/TMAG.2009.2012770