Feedback Analysis in Percutaneous Bone-Conduction Device and Bone-Conduction Implant on a Dry Cranium
Artikel i vetenskaplig tidskrift, 2012

Hypothesis: The bone-conduction implant (BCI) can use a higher gain setting without having feedback problems compared with a percutaneous bone-conduction device (PBCD). Background: The conventional PBCD, today, is a common treatment for patients with conductive hearing loss and single-sided deafness. However, there are minor drawbacks reported related to the percutaneous implant and specifically poor high-frequency gain. The BCI system is designed as an alternative to the percutaneous system because it leaves the skin intact and is less prone to fall into feedback oscillations, thus allowing more high-frequency gain. Methods: Loop gains of the Baha Classic 300 and the BCI were measured in the frequency range of 100 to 10,000 Hz attached to a Skull simulator and a dry cranium. The Baha and the BCI positions were investigated. The devices were adjusted to full-on gain. Results: It was found that the gain headroom using the BCI was generally 0 to 10 dB better at higher frequencies than using the Baha for a given mechanical output. More specifically, if the mechanical output of the devices were normalized at the cochlear level the improvement in gain headroom with the BCI versus the Baha were in the range of 10 to 30 dB. Conclusion: Using a BCI, significantly higher gain setting can be used without feedback problems as compared with using a PBCD.

Gain headroom

Sound radiation

Baha

Implantable transducer

Percutaneous bone-conduction device

Stability

Bone-conduction implant

Feedback

Loop gain

Författare

Hamidreza Taghavi

Chalmers, Signaler och system, Signalbehandling och medicinsk teknik

Bo Håkansson

Chalmers, Signaler och system, Signalbehandling och medicinsk teknik

Sabine Reinfeldt

Chalmers, Signaler och system, Signalbehandling och medicinsk teknik

Måns Eeg-Olofsson

Göteborgs universitet

Shirin Sadat Akhshijan

Chalmers, Signaler och system, Signalbehandling och medicinsk teknik

Otology and Neurotology

1531-7129 (ISSN) 1537-4505 (eISSN)

Vol. 33 3 413-420

Ämneskategorier

Reglerteknik

DOI

10.1097/MAO.0b013e3182487fc8

PubMed

22410731

Mer information

Skapat

2017-10-07