Evolution of an UWB antenna for hyperthermia array applicator
Paper i proceeding, 2012

Designs of three UWB antennas and their suitability to be used as a basis antenna element in the hyperthermia applicator for the treatment of tumors in head and neck region are considered. The antennas with reflection coefficient less than -10 dB at frequency range 350 MHz to 1 GHz are immersed in water bolus. The evaluation of the SAR distributions was carried out by comparing the different cross-sections of muscle equivalent phantom. Numerical simulations were performed in CST Microwave studio and verified experimentally via measurements of S-parameters and SAR distributions in phantoms.

Författare

Hana Dobsicek Trefna

Chalmers, Signaler och system, Signalbehandling och medicinsk teknik

Azeem Imtiaz

Hoi-Shun Lui

Chalmers, Signaler och system, Signalbehandling och medicinsk teknik

Tonny Rubaek

Chalmers, Signaler och system, Signalbehandling och medicinsk teknik

Mikael Persson

Chalmers, Signaler och system, Signalbehandling och medicinsk teknik

Proceedings of 6th European Conference on Antennas and Propagation, EuCAP 2012. Prague, 26-30 March 2012.

1046 - 1048
978-145770918-0 (ISBN)

Ämneskategorier

Annan medicinteknik

Styrkeområden

Livsvetenskaper och teknik (2010-2018)

DOI

10.1109/EuCAP.2012.6206294