The bone-implant interface of dental implants in humans on the atomic scale
Artikel i vetenskaplig tidskrift, 2017

Osseointegration of dental implants occurs on a hierarchy of length scales down to the atomic level. A deeper understanding of the complex processes that take place at the surface of an implant on the smallest scale is of interest for the development of improved biomaterials. To date, transmission electron microscopy (TEM) has been utilized for examination of the bone-implant interface, providing details on the nanometer level. In this study we show that TEM imaging can be complemented with atom probe tomography (APT) to reveal the chemical composition of a Ti-based dental implant in a human jaw on the atomic level of resolution. As the atom probe technique has equal sensitivity for all elements, it allows for 3 dimensional characterizations of osseointegrated interfaces with unprecedented resolution. The APT reconstructions reveal a Ca-enriched zone in the immediate vicinity of the implant surface. A surface oxide of some 5 nm thickness was measured on the titanium implant, with a sub-stoichiometric composition with respect to TiO2. Minor incorporation of Ca into the thin oxide film was also evident. We conclude that the APT technique is capable of revealing chemical information from the bone-implant interface in 3D with unprecedented resolution, thus providing important insights into the mechanisms behind osseointegration. Statement of Significance Osseointegration of dental implants occurs on a hierarchy of length scales down to the atomic level. A deeper understanding of the complex processes that take place at the surface of an implant on the smallest scale is of interest for the development of improved biomaterials. To date, transmission electron microscopy (TEM) has been utilized for examination of the bone-implant interface, providing details on the nanometer level. In this study we show that TEM imaging can be complemented with atom probe tomography (APT) to reveal the chemical composition of a Ti-based dental implant in a human jaw on the atomic level of resolution. Correlative microscopy ensures the accuracy of APT reconstructions and helps provide both chemical and structural information of the bone-implant interface on the smallest of length scales.

Osseointegration

Transmission electron microscopy

Dental implant

Atom probe tomography

Titanium implant

Författare

Gustav Sundell

Chalmers, Fysik, Materialens mikrostruktur

Christer Dahlin

Göteborgs universitet

NU-sjukvården

Martin Andersson

Chalmers, Kemi och kemiteknik, Tillämpad kemi

Mattias Thuvander

Chalmers, Fysik, Materialens mikrostruktur

Acta Biomaterialia

1742-7061 (ISSN) 18787568 (eISSN)

Vol. 48 445-450

Ämneskategorier

Fysik

DOI

10.1016/j.actbio.2016.11.044

Mer information

Senast uppdaterat

2021-01-05