A First Evaluation of the Realistic Supelec-Breast Phantom
Artikel i vetenskaplig tidskrift, 2017

A microwave tomographic system is used to evaluate the realistic breast phantom developed at the Supelec institute. The system utilizes 16 monopole antennas and a modern vector network analyzer (VNA) to measure the phantoms influence on the S-parameters. An iterative algorithm is then used to solve the inverse problem and reconstruct a 2-D plane transecting the phantom. The reconstructed images are compared to the ones recovered from a cylindrical phantom of equivalent phantom media. The results show that both phantoms are possible to reconstruct, although the interior of the Supelec phantom is more challenging.

Tomography

Phantoms

Dielectrics

Microwave imaging

Image reconstruction

Breast cancer