Phase calibration of a software defined radio system for medical applications
Paper i proceeding, 2019

Accurate phase measurement is crucial for many medical applications. We presented the challenge of using a software defined radio system for phase measurement and proposed two different calibration strategies to address this problem. Both strategies were implemented and evaluated. The results show that the first strategy gives better measurement repeatability than the second one. With this strategy, the random phase offset between the transmitter and receiver can be effectively compensated and the obtained phase measurement uncertainty is in the order of sub-degrees.

Software defined radio

Phase measurement

Biomedical applications

Författare

Xuezhi Zeng

Chalmers, Elektroteknik, Signalbehandling och medicinsk teknik

Laura Guerrero Orozco

Chalmers, Elektroteknik, Signalbehandling och medicinsk teknik

Andreas Fhager

Chalmers, Elektroteknik, Signalbehandling och medicinsk teknik

Asia-Pacific Microwave Conference Proceedings, APMC

Vol. 2019-December 1414-1416 9038268
978-172813517-5 (ISBN)

2019 IEEE Asia-Pacific Microwave Conference, APMC 2019
Singapore, Singapore,

Ämneskategorier

Signalbehandling

DOI

10.1109/APMC46564.2019.9038268

Mer information

Senast uppdaterat

2024-01-03