Biomechanical Analysis of Body Movements of Myoelectric Prosthesis Users During Standardized Clinical Tests
Artikel i vetenskaplig tidskrift, 2023

Objective: The objective clinical evaluation of user's capabilities to handle their prosthesis is done using various tests which primarily focus on the task completion speed and do not explicitly account for the potential presence of compensatory motions. Given that the excessive body compensation is a common indicator of inadequate prosthesis control, tests which include subjective observations on the quality of performed motions have been introduced. However, these metrics are then influenced by the examiner's opinions, skills, and training making them harder to standardize across patient pools and compare across different prosthetic technologies. Here we aim to objectively quantify the severity of body compensations present in myoelectric prosthetic hand users and evaluate the extent to which traditional objective clinical scores are still able to capture them. Methods: We have instructed 9 below-elbow prosthesis users and 9 able-bodied participants to complete three established objective clinical tests: Box-and-Blocks-Test, Clothespin-Relocation-Test, and Southampton-Hand-Assessment-Procedure. During all tests, upper-body kinematics has been recorded. Results: While the analysis showed that there are some correlations between the achieved clinical scores and the individual body segment travel distances and average speeds, there were only weak correlations between the clinical scores and the observed ranges of motion. At the same time, the compensations were observed in all prosthesis users and, for the most part, they were substantial across the tests. Conclusion: The sole reliance on the currently available objective clinical assessment methods seems inadequate as the compensatory movements are prominent in prosthesis users and yet not sufficiently accounted for.

Prosthetics

box and blocks

Task analysis

Wrist

body compensation

bionic hand

Force

Time measurement

CPRT

Biomechanics

SHAP

myocontrol

Prosthetics

motion capture

Junctions

Författare

Ivan Vujaklija

Aalto-Yliopisto

Moon Ki Jung

Imperial College London

Timothy Hasenoehrl

Medizinische Universität Wien

Aidan D. Roche

St John's Hospital

Agnes Sturma

Imperial College London

Silvia Muceli

Chalmers, Elektroteknik, Signalbehandling och medicinsk teknik

Richard Crevenna

Medizinische Universität Wien

Oskar C. Aszmann

Medizinische Universität Wien

Dario Farina

Imperial College London

IEEE Transactions on Biomedical Engineering

0018-9294 (ISSN) 15582531 (eISSN)

Vol. 70 3 789-799

Ämneskategorier

Medicinsk bioteknologi

DOI

10.1109/TBME.2022.3202250

PubMed

36037457

Mer information

Senast uppdaterat

2023-03-15