Study on the sensitivity of image reconstruction to the measurement uncertainty in microwave tomography
Paper in proceeding, 2010

We investigated the effects of measurement uncertainty on the image reconstructions for different breast models. Three different contrast levels between the background tissue and tumor were chosen. It was shown that the reconstruction of lower contrast breast model was more sensitive to the measurement uncertainty. For a breast model with contrast as low as 1.1 : 1, an amplitude uncertainty of 0.2 dB or a phase uncertainty of 2 degrees would result in an unsuccessful reconstruction.

Author

Xuezhi Zeng

Chalmers, Signals and Systems, Signal Processing and Biomedical Engineering

Andreas Fhager

Chalmers, Signals and Systems, Signal Processing and Biomedical Engineering

Mikael Persson

Chalmers, Signals and Systems, Signal Processing and Biomedical Engineering

EuCAP 2010 - The 4th European Conference on Antennas and Propagation

5505403
978-847653472-4 (ISBN)

Subject Categories

Electrical Engineering, Electronic Engineering, Information Engineering

Other Electrical Engineering, Electronic Engineering, Information Engineering

ISBN

978-847653472-4

More information

Created

10/8/2017