Study on the sensitivity of image reconstruction to the measurement uncertainty in microwave tomography
Paper i proceeding, 2010

We investigated the effects of measurement uncertainty on the image reconstructions for different breast models. Three different contrast levels between the background tissue and tumor were chosen. It was shown that the reconstruction of lower contrast breast model was more sensitive to the measurement uncertainty. For a breast model with contrast as low as 1.1 : 1, an amplitude uncertainty of 0.2 dB or a phase uncertainty of 2 degrees would result in an unsuccessful reconstruction.

Författare

Xuezhi Zeng

Chalmers, Signaler och system, Signalbehandling och medicinsk teknik

Andreas Fhager

Chalmers, Signaler och system, Signalbehandling och medicinsk teknik

Mikael Persson

Chalmers, Signaler och system, Signalbehandling och medicinsk teknik

EuCAP 2010 - The 4th European Conference on Antennas and Propagation

5505403
978-847653472-4 (ISBN)

Ämneskategorier

Elektroteknik och elektronik

Annan elektroteknik och elektronik

ISBN

978-847653472-4

Mer information

Skapat

2017-10-08