Accuracy investigation of an ultra-wideband time domain microwave imaging system
Paper in proceeding, 2011

We investigate the measurement accuracy of an ultra-wideband time domain microwave tomography system. A phantom is imaged by means of the system and the reconstructed images are compared to those obtained with an ultra-wideband frequency domain system. The results suggest that the permittivity reconstructions obtained by means of the time domain and the frequency domain systems are very similar to each other, but the conductivity profile reconstructed from time domain measurements presents worse quality than that obtained from frequency domain measurements.

Author

Xuezhi Zeng

Chalmers, Signals and Systems, Signal Processing and Biomedical Engineering

Andreas Fhager

Chalmers, Signals and Systems, Signal Processing and Biomedical Engineering

Peter Linner

Chalmers, Microtechnology and Nanoscience (MC2), Microwave Electronics

Mikael Persson

Chalmers, Signals and Systems, Signal Processing and Biomedical Engineering

Herbert Zirath

Chalmers, Microtechnology and Nanoscience (MC2), Microwave Electronics

Proceedings of the 5th European Conference on Antennas and Propagation, EUCAP 2011. Rome, 11-15 April 2011

1928-1932
978-888202074-3 (ISBN)

Areas of Advance

Life Science Engineering (2010-2018)

Subject Categories

Electrical Engineering, Electronic Engineering, Information Engineering

ISBN

978-888202074-3

More information

Created

10/7/2017