Development of a Time Domain Microwave System for Medical Diagnostics
Journal article, 2014

In this paper, a time-domain system dedicated to medical diagnostics has been designed, a prototype has been built and its performance has been evaluated. Measurements show that the system has a 3-dB bandwith of about 3.5 GHz and a signal to noise ratio over 40 dB in the frequency range about 800 MHz to 3.8 GHz. The system has been used to perform a microwave tomographic image reconstruction test. The same target was reconstructed based on data measured with a network analyzer. A comparison between the images shows very small differences, and proves the functionality of the time domain system.

Microwave imaging

time domain measurements

ultrawideband (UWB)

Author

Xuezhi Zeng

Chalmers, Signals and Systems, Signal Processing and Biomedical Engineering

Andreas Fhager

Chalmers, Signals and Systems, Signal Processing and Biomedical Engineering

Zhongxia Simon He

Chalmers, Microtechnology and Nanoscience (MC2), Microwave Electronics

Mikael Persson

Chalmers, Signals and Systems, Signal Processing and Biomedical Engineering

Peter Linner

Chalmers, Microtechnology and Nanoscience (MC2), Microwave Electronics

Herbert Zirath

Chalmers, Microtechnology and Nanoscience (MC2), Microwave Electronics

IEEE Transactions on Instrumentation and Measurement

0018-9456 (ISSN) 1557-9662 (eISSN)

Vol. 63 12 2931-2939 6840331

Areas of Advance

Information and Communication Technology

Subject Categories

Other Electrical Engineering, Electronic Engineering, Information Engineering

DOI

10.1109/TIM.2014.2326277

More information

Latest update

4/5/2022 6