A laboratory exercise on systematic effects in gamma ray spectrometry
Artikel i vetenskaplig tidskrift, 2015

A laboratory exercise for calculation of true coincidence summing correction factors as well as calculating the effect of deviations between sample and standard source (filling height) was developed. This laboratory exercise was held in a masters course in nuclear chemistry the first time during fall 2013. The aim of the exercise was to high-light the importance of correcting for biases due to different systematic effects in gamma spectrometric measurements.

Systematic effects

Gamma spectrometry

True coincidence summing

EFFTRAN.

Författare

Henrik Ramebäck

Chalmers, Kemi och kemiteknik, Energi och material

S. Jonsson

Totalförsvarets forskningsinstitut (FOI)

Stefan Allard

Chalmers, Kemi och kemiteknik, Energi och material

Christian Ekberg

Chalmers, Kemi och kemiteknik, Energi och material

T. Vidmar

Belgian Nuclear Research Center (SCK CEN)

Journal of Radioanalytical and Nuclear Chemistry

0236-5731 (ISSN) 1588-2780 (eISSN)

Vol. 304 1 467-471

DOI

10.1007/s10967-014-3525-z

Mer information

Senast uppdaterat

2018-02-27